Redefining Diabetes Care: Inferring Blood Glucose Level from Breath
Acknowledgments
This project is a collaboration between the Center for Applied AI and the following researchers:
Dr. Yuan Wen (Institute for Biomedical Informatics, Department of Physiology)
Dr. Laura Brown (Department of Physiology)
Dmitry “Dima” Strakovsky (College of Fine Arts)
Background
Volatile organic compounds (VOCs) are chemicals that can easily evaporate at room temperature and are commonly found in various products and materials such as paints, cleaning supplies, fuels, and solvents. The measurement of these compounds has been evolving over decades. Today, thanks to advancements in collection techniques, collection devices fit in the palm of your hand and can detect trace amounts of VOCs. These devices are used for many applications, but we are interested in their applications to healthcare. Specifically, can we collect breath samples from a patient to predict blood glucose levels using a machine-learning model? The VOC sensor device we use is named the “Smell Inspector” and is described in the “Smell Inspector” section below.
We aim to 1) create a dataset with blood glucose measurements and 2) infer the blood glucose measurement from a breath sample in real time.
Introduction
This project is in the early phases. However, we have begun to collect breath samples from 4 volunteers in our office. None of these volunteers have diabetes and do not have access to a measurement device for blood glucose, so we decided to test the effectiveness of the VOC sensor by classifying peppermint breath vs. normal breath. By doing this, we can see how sensitive the smell inspector is when classifying something aromatic while testing various machine learning models. The experimental setup and results are in the “Peppermint vs. Normal Breath” section below. These measurements will not be included in the final dataset or model; they simply help demonstrate the best setup for the actual blood glucose project.
Smell Inspector
The device used to collect these samples is the “Smell Inspector, ” developed by SmartNanotubes Technologies in Germany (found here). Figure 1 shows the smell inspector with an additional mounting plate designed in-house to connect to a breath chamber.
The smell sensor collects and sends data at approximately 2 seconds increments. This resolution is not ideal for breath samples, as many can only exhale for 7-8 seconds. Based on the results below, it may not matter that we are collecting a relatively low number of samples per exhale.
Collection Chamber
The collection chamber was built by Dima Strakovsky and has been a great starting point for breath collection. The chamber is completely modular, where we can add or remove vents and change the distance from the mouthpiece to the sensor inlet. Figure 2-5 shows the chamber with all panels removed (except for the bottom panel) and the mouthpiece directly facing the smell inspector.
The smell inspector can move to any distance within the collection chamber simply by pushing the device farther into the chamber. The inspector creates a seal with the chamber by using a custom mounting plate that includes rubber grommets that hold the device in place without needing screws or other hardware. This can be seen in Figure 2. By allowing the device to move in and out of the chamber, we were able to collect samples at three places: short (sensor placed ~1 inch from the mouthpiece), medium (sensor placed in the middle of the chamber), and long (sensor placed at the end of chamber). Doing this allowed us to discover which distance from the mouthpiece provided us with the best results.
Mouthpiece
Each mouthpiece is single-use only and comes in a variety of designs. In Figure 4, the two blue mouthpieces include a particle filter that claims to stop viruses and excessive moisture but allows VOCs to pass through. Additionally, there is also a direct connection mouthpiece without a filter. There is not a large difference between moisture entering the chamber, and it did not appear that using one over the other made any difference when it came to measurement when comparing the filtered vs non-filtered version. However, using the filtered mouthpiece would be recommended to prevent excessive amounts of saliva and other large particles from entering the chamber.
Peppermint Breath vs. Normal Breath
As mentioned earlier, we took measurements from 3 sensor positions within the chamber. Each volunteer exhaled into the chamber three times at each of these positions. This created nine measurements for each volunteer for “Peppermint Breath” and nine measurements for “Normal Breath”. After each exhalation, the smell inspector was removed from the chamber, and both the inspector and the chamber were wafted around for at least 1 minute before placing the inspector back in the chamber for the subsequent measurement. This way, the chamber could be cleared of a high concentration of VOCs to re-establish a baseline.
Normal Collection: Volunteers were told to stop eating and drinking for one hour before collecting breath samples.
Peppermint Collection: Volunteers were told to stop eating and drinking for one hour before collecting breath samples. Volunteers were then given a peppermint to eat at each sensor distance (one at short, one at medium, one at long). After finishing the Peppermint, volunteers were asked to wait one minute and then exhale into the chamber.
Results
Of the four volunteers, each produced about 120 samples over nine measurements. More data is needed to train a robust machine-learning model. Therefore, we combined the samples for all volunteers at each distance.
We used CLASSify to train each dataset with seven models: Random Forest (RF), XGBoost (XG), Gradient Boosting (GB), Histogram Gradient Boosting (HGB), Bagging (B), SGDClassifier (SGD), Logistic Regression (LR), and K-Nearest Neighbors (KNN). Below are the results for each dataset and the best-performing model. If multiple models are listed in the best model column, each of these models performed exactly the same. Each dataset is split into 80% for training and 20% for testing. The classes (peppermint vs normal) are almost perfectly balanced in each dataset.
Sensor Distance | # Samples | Best Model | AUC | Accuracy | Sensitivity | Specificity |
---|---|---|---|---|---|---|
Short | 262 | SGD | 1.0 | 1.0 | 1.0 | 1.0 |
Medium | 262 | GB | 1.0 | 1.0 | 1.0 | 1.0 |
Long | 269 | XG, B | 1.0 | 1.0 | 1.0 | 1.0 |
Short_Medium_Long_Combined | 793 | HGB, SGD | 1.0 | 1.0 | 1.0 | 1.0 |
Based on this table, the models are probably overfitting the data since there is little data to train on right now. On the other hand, since these are binary classification models, the introduction of peppermint might be strong enough to clearly separate normal breath from peppermint breath. Each model works perfectly. In the following weeks, we will also be adding two more classes for butterscotch and strawberry-flavored candy and re-train the model to see how the results change.
Future updates will be posted here.